• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT225
Device Problem Optical Decentration (1360)
Patient Problems Vitrectomy (2643); No Code Available (3191)
Event Date 08/09/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that a zct225 21.5 diopter intraocular lens (iol) was implanted into the right eye (od) on (b)(6)2017.It was then explanted the next day, (b)(6)2017, because of poor centration and poor vision.Pre-operative visual acuity was 20/200 and post-operative visual acuity was also 20/200.There was an incision enlargement and vitrectomy performed.The patient's vision improved after surgery with a post-operative visual acuity of 20/50.The replacement lens was the same diopter but a different model, a za9003 lens.No additional information was provided.
 
Manufacturer Narrative
Device evaluation: product testing could not be performed because the product was not returned.The reported complaint was not verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The dfu provide the customer with proper usage instructions and guidelines.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Type of Device
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
van swietenlaan 5
groningen 9728 NX
NL   9728 NX
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key6843877
MDR Text Key84933516
Report Number9614546-2017-00838
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474538788
UDI-Public(01)05050474538788(17)200226
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date02/26/2020
Device Model NumberZCT225
Device Catalogue NumberZCT2250215
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-