| Model Number |
205113 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/11/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event has been determined to be a post-placement/pre-load implant failure.This implant loss suggests that the source of the problem was likely to stem from procedural errors and/or the lack of osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, or patient behavior.Proper intended use of the implant is described in its instructions for use.Physical analysis was not performed.
|
| |
|
Event or Problem Description
|
|
Time of loss in relation to the implantation was up to 1 yr after prosthetic restoration.Bone quality at the time of implant failure type iv.Primary stability and osseointegration was not achieved.Augmentation procedure was not performed at time of surgery.Patient has diabetes and xerostomia.
|
| |
|