| Model Number |
LF1520 |
| Medical Device Problem Code |
Energy Output To Patient Tissue Incorrect (1209)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample was requested but to date has not been received for evaluation.If the sample or additional information pertinent to the incident is obtained, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a low anterior resection, the device did not seal tissue.The device activated and an end tone was heard indicating a completed sealing cycle.There was no patient injury, and a new device was used to continue the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one used lf1520 ligasure device was received, and evaluation of the returned sample could not confirm the reported issue.Visual inspection found no defects.The device was activated multiple times on porcine kidney samples, pressing the button in various locations on a range of vessel sizes to detect any activation issues.All seal cycles were completed satisfactorily, and a visual seal effect with an end tone was observed for each application.The investigation found the device to function normally and within specifications.There are factors during use that can affect seal quality, and the instructions for use included with this product lists measures to ensure sealing effectiveness.If information is provided in the future, a supplemental report will be issued.
|
| |
|