• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COSMAN MEDICAL NITINOL TC ELECTRODE; ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COSMAN MEDICAL NITINOL TC ELECTRODE; ELECTRODES Back to Search Results
Model Number TCN-10
Device Problems Break (1069); Crack (1135); Material Discolored (1170); Contamination /Decontamination Problem (2895)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/21/2017
Event Type  malfunction  
Manufacturer Narrative
Additional suspect medical device components involved: model #: tcn-10 lot #: 050914, 100214, 020916, and 021816 description: nitinol tc electrode, 100 mm; model #: tcn-15 lot #: 062414 and 122815 description: nitinol tc electrode, 150 mm.
 
Event Description
Device evaluation performed on the returned electrodes showed that electrode lot # 050914 had discolored epoxy with a chip out.Electrode with lot #100214 had discolored epoxy and was cracked.Electrodes with lot # 123115, 021816, 062414, 020916, and 122815 are discolored and have a chip out.If the epoxy is subjected to too many autoclave cycles, the epoxy becomes brittle and discolored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITINOL TC ELECTRODE
Type of Device
ELECTRODES
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key6845620
MDR Text Key85003860
Report Number3006630150-2017-03272
Device Sequence Number1
Product Code GXI
UDI-Device Identifier00813250013054
UDI-Public00813250013054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberTCN-10
Device Lot Number123115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-