Additional suspect medical device components involved: model #: tcn-10 lot #: 050914, 100214, 020916, and 021816 description: nitinol tc electrode, 100 mm; model #: tcn-15 lot #: 062414 and 122815 description: nitinol tc electrode, 150 mm.
|
Device evaluation performed on the returned electrodes showed that electrode lot # 050914 had discolored epoxy with a chip out.Electrode with lot #100214 had discolored epoxy and was cracked.Electrodes with lot # 123115, 021816, 062414, 020916, and 122815 are discolored and have a chip out.If the epoxy is subjected to too many autoclave cycles, the epoxy becomes brittle and discolored.
|