Siemens requested the patient sample and information on the diagnosis and treatment of the patient.Additional information and the sample were not provided.Siemens service went on site and inspected and tested the advia centaur xp in diagnostics.The system checks were good and within limits.The acid and base dispense properly.A precision study of afp, ca 15-3 and cea showed no issues.A precision study of the sample in question was performed and no issues were observed.Qc passed within limits.Siemens reviewed the customer's calibrations and control values and did not observe any issues.The root cause of the elevated result with this one sample cannot be determined.The sample was not tested on any other lot of reagent.The expected results section of the instructions for use states: "in this study, 98.4% of the apparently healthy subjects had afp values less than 8.1 ng/ml." the table shows that 12 apparently healthy subjects had values between 8.1 ng/ml and 20.0 ng/ml.The warning in the intended use section of the instructions for use states: "the concentration of afp in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity.The results reported by the laboratory to the physician must include the identity of the afp assay used.Values obtained with different afp assay methods cannot be used interchangeably." the interpretation of results section of the instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." based on the information available, advia centaur afp lot 188 is performing as intended.No further investigation required.
|
|
Customer observed a reproducibly, elevated advia centaur xp alpha-fetoprotein (afp) result on one patient sample compared to historical results and an alternate method.The physician questioned the result since historically, the result was lower.There are no reports that treatment was altered or prescribed or adverse health consequences due to the elevated advia centaur xp afp results.
|