| Model Number |
EPP3CR |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Lead(s), Burn(s) From (3161)
|
| Date of Event |
07/20/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the supplier¿s results of investigation, all materials, production processes and inspections met the specification requirements for the neonatal electrode code epp3cr.The device history record review did not indicate any exception that could lead to the reported incident.The supplier investigated and conducted clinical tests and the results did not reveal any negative skin issues.This is the first complaint of this type for this product.From the investigation the root cause could not be determined.No additional action will be taken at this time, but we will continue to monitor the trend for this type of incident.
|
| |
|
Event or Problem Description
|
|
User facility reported a possible burn from neonatal leads.Leads were changed 24 hours prior to event date (b)(6) 2017.At time of event bedside nurse changed leads to the neolead micro electrodes.Green lead was affected and changed.Patient site reviewed and left mid abdomen found to have an intact flat blister circular 1 cm x 1 cm area with a faint erythematous circular region the size of the electrode.Flat blister appears intact with scant dry exudate.
|
| |
|