| Model Number |
MMT-7008B |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 390 mg/dl and the sensor glucose was 90 mg/dl at the time of the incident.Customer reports sensors are only lasting 2-3 days, around day 3 they start to show sensor glucose value of 100 mg/dl no matter what blood glucose is.In many cases customer states that it will put him in threshold suspend.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.Customer was assisted with troubleshooting.The sensor will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened/used sensor and performed continuity resistance test.Found sensor passed per specifications and performed the sensor initialization test using a new lab transmitter with a 7xx/5xx paradigm insulin pump.Connected the sensor to the transmitter and placed it in 100 bts solution, confirmed the communication icon was displayed and that the transmitter was flashing while connected to the sensor.The sensor functioned properly.Cannot performed bts test as the sensor is beyond its expiration date.
|
| |
|