|
As reported, patient had picc line placed (b)(6), 2017.Line became non-functioning.Picc line service nurse removed picc line from patient on (b)(6) 2017 and found it to contain a segment which appeared to be "deflated and stretched." another picc line was placed.No sample is expected to be returned.
|
|
A review of the device history records was performed for the indicated packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly and performance specifications.The august 2017 angiodynamics complaint report was reviewed for the bioflo picc product family and the failure mode "device/catheter malfunctioned - removed." no adverse trend was identified.The end user's reported complaint description of "picc appeared deflated and stretched" is not confirmed, nor can a root cause be determined, as no sample was returned for evaluation.Angiodynamics' incoming inspection of the catheter tubing includes a visual inspection based on an aql, which includes but is not limited to, visualizing for dents in the catheter tubing.Dents that can be felt, but not seen are acceptable; dents that can be felt and seen are rejectable.In addition, a guidewire insertion test is performed.The guidewire must pass through the entire length of the extrusion with no resistance.Manufacturing process controls include visual inspection for damage or defects.All catheters are subjected to a guidewire insertion test to verify lumen patency.The guidewire must pass through the catheter with little to no resistance to be considered acceptable.In addition, all catheters are 100% sprint tested after the guidewire verification to ensure the catheter does not leak.The directions for use provided with the picc device provide guidance on catheter placement, and flushing.((b)(4)).
|