Model Number ZXR00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Dry Eye(s) (1814); Visual Impairment (2138); No Code Available (3191)
Date of Event 08/14/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxr00 25.5 diopter intraocular lens was explanted from the patient's left eye on (b)(6)2017 and replaced with a new lens.There was no incision enlargement, vitrectomy, or sutures used.No additional information was provided.
 
Additional Manufacturer Narrative
Additional information received reported the reason for explant was because of poor vision related to dry eyes.Also, the name of the doctor was dr.(b)(6).Reportedly, the patient is doing well after surgery.No additional information was provided.Device available for evaluation?: yes, returned to manufacturer on 08/24/2017.Initial reporter name: dr.(b)(6).Health professional: yes.Occupation: physician.Device returned to manufacturer?: yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye showed the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key6850342
Report Number9614546-2017-00846
Device Sequence Number18849079
Product Code POE
UDI-Device Identifier05050474579200
UDI-Public(01)05050474579200(17)220606
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date06/06/2022
Device Model NumberZXR00
Device Catalogue NumberZXR00U0255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2017
Initial Date Received by Manufacturer 08/14/2017
Supplement Date Received by Manufacturer10/02/2017
Initial Report FDA Received Date09/07/2017
Supplement Report FDA Received Date10/26/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-