| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Dry Eye(s) (1814); Visual Impairment (2138); No Code Available (3191)
|
| Date of Event |
08/14/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxr00 25.5 diopter intraocular lens was explanted from the patient's left eye on (b)(6)2017 and replaced with a new lens.There was no incision enlargement, vitrectomy, or sutures used.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reported the reason for explant was because of poor vision related to dry eyes.Also, the name of the doctor was dr.(b)(6).Reportedly, the patient is doing well after surgery.No additional information was provided.Device available for evaluation?: yes, returned to manufacturer on 08/24/2017.Initial reporter name: dr.(b)(6).Health professional: yes.Occupation: physician.Device returned to manufacturer?: yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye showed the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|