Catalog Number 1012268-12U
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual and functional inspections were performed on the returned device.The reported crack and leak were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported crack and leak appear to be related to operation context.There is no indication of a product quality issue with respects to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during device preparation and prior to use in the anatomy, the trek leaked at the hub area due to a crack in the luer.There was no patient involvement; the device was not used.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK CORONARY DILATATION CATHETER
Common Device Name
BALLOON DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6850590
Report Number2024168-2017-07346
Device Sequence Number488597
Product Code LOX
UDI-Device Identifier08717648176555
UDI-Public08717648176555
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K110617
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 09/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number1012268-12U
Device Lot Number70605G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2017
Initial Report FDA Received Date09/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-