• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS / ANGIOVAC; CATHETER, CANNULA & TUBING, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS / ANGIOVAC; CATHETER, CANNULA & TUBING, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number H965251850
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device from the reported event is expected to be returned to angiodynamics but has not yet arrived.Upon receipt of the device and completion of the investigation, a supplemental medwatch will be submitted.((b)(4)).
 
Event or Problem Description
As reported end user hospital noted a small hole in the packaging of the angiovac cannula, thereby compromising the sterility.The device was not used on a patient.It will be returned to angiodynamics for evaluation.
 
Additional Manufacturer Narrative
A review of the device history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.The angiodynamics complaint report was reviewed for the angiovac product family and the failure mode,"package - hole/perforated." no adverse trends were identified.As no sample was returned for evaluation, the exact root cause of the reported issue is unable to be determined.All pouches are 100% inspected per angiodynamics procedures during the sealing and final box processes.Employees involved in the packaging/final boxing of the reported lot of product have been made aware of this complaint.The directions for use (dfu) packaged with the angiovac contain the following warning: "contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged." ((b)(4)).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS / ANGIOVAC
Common Device Name
CATHETER, CANNULA & TUBING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key6851211
Report Number1317056-2017-00073
Device Sequence Number518797
Product Code DWF
UDI-Device IdentifierH965251850
UDI-PublicH965251850
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K142593
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/31/2017
Device Catalogue NumberH965251850
Device Lot Number4943462
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2017
Supplement Date Received by Manufacturer08/18/2017
Initial Report FDA Received Date09/08/2017
Supplement Report FDA Received Date10/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-