• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT MEDICAL SHILEY; TUBE, BRONCHIAL (W/WO CONNECTOR) Back to Search Results
Model Number 125041
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report the cuff on the endo bronchial tube would not deflate.Customer indicated that there was no patient involvement associated to this event.
 
Additional Manufacturer Narrative
Evaluation summary: the sample was returned and visual examination showed that the shape returned unit had been altered using the stylet wire.The stylet wire was removed and inflation/deflation testing was successful.There was no fault found with the device.The probable root cause is that the cuff test was carried out when the shape of the tube was incorrectly altered with the stylet wire.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, BRONCHIAL (W/WO CONNECTOR)
Manufacturer (Section D)
MALLINCKRODT MEDICAL
cornamaddy
athlone 3810
MDR Report Key6851732
Report Number8020889-2017-05130
Device Sequence Number13577681
Product Code BTS
UDI-Device Identifier30884522008765
UDI-Public30884522008765
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K771219
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2021
Device Model Number125041
Device Catalogue Number125041
Device Lot Number201607016X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2018
Initial Date Received by Manufacturer 08/16/2017
Supplement Date Received by Manufacturer03/05/2018
Initial Report FDA Received Date09/08/2017
Supplement Report FDA Received Date04/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-