Model Number 236-02-108
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 08/08/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient experiencing mal-alignment.Stryker products were used.The patella wasn't removed.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to the patient experiencing a mal-alignment.The in-vivo length of patient service for the implant was 9.2 months.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed as non-product related.The root cause of this event was the patient had a mal-positioned knee implant and required a revision to correct.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3DKNEE
Common Device Name
FEMUR, POROUS COATED, 3D, SIZE8, RT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key6853547
Report Number1644408-2017-00786
Device Sequence Number10779624
Product Code MBH
UDI-Device Identifier00888912111812
UDI-Public(01)00888912111812
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K032905
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2021
Device Model Number236-02-108
Device Catalogue Number236-02-108
Device Lot Number972A1099
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/08/2017
Supplement Date Received by Manufacturer09/29/2017
Initial Report FDA Received Date09/08/2017
Supplement Report FDA Received Date10/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age83 YR
-
-