| Model Number |
TCN-15 |
| Medical Device Problem Codes |
Break (1069); Material Discolored (1170); Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/25/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device analysis performed showed that the epoxy has chip out and discoloration.The color of newly cured epoxy is amber, light brown.If the epoxy is subjected to too many autoclave cycles it becomes brittle and discolored.
|
| |
|
Search Alerts/Recalls
|