• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; STEM, HUMERAL SIZE6 RSP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; STEM, HUMERAL SIZE6 RSP Back to Search Results
Model Number 506-00-006
Health Effect - Clinical Code Fall (1848)
Date of Event 08/10/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - the surgeon removed the stem and insert.The patient fell and broke bone, fracturing the stem.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to the patient having a fractured bone and stem.The in-vivo length of patient service for the implant was 4.3 years.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The complaint investigation history was reviewed and no trends or on-going issues were deemed to be present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had a fractured bone and stem from a fall.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
STEM, HUMERAL SIZE6 RSP
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key6855004
Report Number1644408-2017-00755
Device Sequence Number16267329
Product Code KWS
UDI-Device Identifier00888912024372
UDI-Public(01)00888912024372
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K041066
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/28/2019
Device Model Number506-00-006
Device Catalogue Number506-00-006
Device Lot Number810C1089
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/10/2017
Supplement Date Received by Manufacturer09/27/2017
Initial Report FDA Received Date09/11/2017
Supplement Report FDA Received Date10/26/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-01-032,LOT 859C1011
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
-
-