| Model Number |
ZM-520PA |
| Medical Device Problem Codes |
Overheating of Device (1437); Smoking (1585); Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the transmitter overheated and was smoking.The battery contacts were burned and melted.There was no other physical damage or fluid intrusion reported.The device was in use on a patient, however no patient injury was reported.They were provided with an exchanged device and sent the failed device in to nihon kohden.It is currently awaiting evaluation.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the transmitter overheated and was smoking.The battery contacts were burned and melted.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the transmitter overheated and was smoking.The battery contacts were burned and melted.There was no other physical damage or fluid intrusion reported.The device was in use on a patient, however no patient injury was reported.They were provided with an exchanged device and sent the failed device in to nihon kohden for evaluation.It was determined in the evaluation that the cause of the unit overheating was due to the customer incorrectly installing the battery in to the unit.
|
| |
|