| Model Number |
ZM-541PA |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Invalid Sensing (2293); Device Operational Issue (2914)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the transmitter was giving inaccurate nibp readings.He tried new hoses and cuffs, as well as running it on a simulator, however the issue continued to persist.The device was in use on a patient, but no patient harm was reported.The biomedical engineer was provided with an exchanged transmitter.He sent the failed device in for repair and it is currently awaiting evaluation.Nihon kohden continues to investigate the reported event.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the transmitter was giving inaccurate nibp readings.
|
| |
|