| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a power (intermittent power) issue.There was no indication that the product caused or contributed to an adverse event.This issue is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 08-nov-2017 with the following findings: a review of the black box showed multiple unexplained power on reset events.The battery compartment was undamaged.The battery cap coin slot was damaged.The battery cap was able to fit securely but was difficult to be removed.The battery cap measurements were within specifications.The battery cap was fastened and then unscrewed a half turn with no reboots occurring.The pump cover was removed to find no intermittent conditions to the power printed circuit board.The intermittent power complaint was unable to be duplicated during investigation.
|
| |
|