| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a call service alarm (call service alarm issue) issue.It was reported that there was a 069 call service alarm.There was no indication that this issue lead to an adverse event.This is being reported because the issue may result in a long term cessation of insulin delivery if the user is unable to resolve the alarm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 08-nov-2017 with the following findings: during investigation, the pump was powered on and a hard call service alarm was shown and could not be reset.The failure is defined as a language file corruption while retrieving the language text.A review of the pump¿s black box and alarm history showed a call service alarm, which is written in the pump's black box.The alarm was found to be caused by a flash chip.Unrelated to the original complaint, the battery compartment was found to be cracked from the case seal traveling to the grip pad.
|
| |
|