| Catalog Number |
THS-LX-C |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/02/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the turbohawk device was visually inspected - biological debris were noted within the laser drilled coil.No damage to the turbohawk unit was noted.The cutter assembly was retracted back into the cutter window.Unknown clear debris was noted behind the cutter.Functional testing was performed - the thumb-switch was attempted to be pushed forward but encountered resistance.A cutter driver from the lab was connected to the turbohawk and activated.The cutter was able to be advanced approximately 2.8cm from the cutter window.The cutter was retracted and observed a strip of clear material wound up around the cutter blank.The strip of material was removed using a tweezers and soaked in water.After approximately 48 hours of soaking the debris did not dissolve.The material was clear and plastic-like.It should be noted the outer layer (hydrophilic coating) on the outer distal rim of the cutter window showed peeling.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician intended to use a turbohawk directional atherectomy device to treat a lesion in the sfa with 70% stenosis.The lesion exhibited severe calcification and tortuosity.It was reported that during the procedure, physician was unable to open the cutter head.Another device was used to complete the procedure.No patient injury reported.
|
| |
|
Search Alerts/Recalls
|