ENCORE MEDICAL L.P. FMP COCR (COBALT CHROMIUM) HEADS; HEAD, FEMORAL, COCR 36MM Back to Search Results
Model Number 497-36-000
Health Effect - Clinical Codes Joint Dislocation (2374); No Code Available (3191)
Date of Event 08/15/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Second revision surgery - due to dislocation.
 
Additional Manufacturer Narrative
The reason for this revision surgery was dislocation.The in-vivo length of patient service for the implant was 26 days.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.This event is deemed to be non-product related.The root cause of this complaint was a revision surgery due to the dislocation.There are multiple factors that may contribute to the event that are outside the control of djo surgical are patient activities, loose joint from inadequate soft tissue, patient bone deterioration or trauma.Due to the short time between the previous and revision surgery, it is also possible that the event may occurred due to improper implant selection/surgical procedure or patient non-compliance with medical instructions.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP COCR (COBALT CHROMIUM) HEADS
Common Device Name
HEAD, FEMORAL, COCR 36MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6861927
Report Number1644408-2017-00771
Device Sequence Number1365043
Product Code JDI
UDI-Device Identifier00888912024051
UDI-Public(01)00888912024051
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/11/2023
Device Model Number497-36-000
Device Catalogue Number497-36-000
Device Lot Number638C1210
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/15/2017
Supplement Date Received by Manufacturer10/06/2017
Initial Report FDA Received Date09/13/2017
Supplement Report FDA Received Date11/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
010-55-025,LOT 006A1313; 010-55-050,LOT 014A1116; 411-00-105,LOT 904B1017; 430-93-062,LOT 047H1010; 932-36-760,LOT 905N1922
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-