• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERTIFLEX, INC. SUPERION INDIRECT DECOMPRESSION SYSTEM; INTERSPINOUS SPACER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERTIFLEX, INC. SUPERION INDIRECT DECOMPRESSION SYSTEM; INTERSPINOUS SPACER Back to Search Results
Model Number 14MM
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Unintended Movement (3026)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Tissue Damage (2104)
Date of Event 08/02/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial report provided by the physician, as conveyed by manufacturer's sales representative, suggested that in post-operative films acquired approximately 100 days following the initial implantation of two superion devices there appeared to be evidence of dislodgement of the superior (l3/l4 device).The product remained in situ, and at the time there was no indication that this was a reportable event.In a telephone conversation between this reporter and the treating physician on (b)(6), the x-rays images were reviewed and it was agreed that the "dislodgement" originally observed by the physician was actually evidence of a fracture of the l4 spinous process.As a result of the determination that an injury (fracture) had occurred, the report was now considered a reportable event.Spinous process fracture is a recognized, low-level risk associated with the superion product, and its occurrence and certain risk factors associated with its likelihood are described in the labeling.
 
Event or Problem Description
Manufacturer's case coverage specialist forwarded x-ray images suggesting "dislodgement of l3-4 superion." device remains in situ.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Common Device Name
INTERSPINOUS SPACER
Manufacturer (Section D)
VERTIFLEX, INC.
2714 loker avenue west
suite 100
carlsbad CA 92010
MDR Report Key6863372
Report Number3005882106-2017-00005
Device Sequence Number633434
Product Code NQO
UDI-Device Identifier00884662000284
UDI-Public00884662000284
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P140004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number14MM
Device Catalogue Number100-9814
Device Lot Number112166R
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2017
Initial Report FDA Received Date09/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-