| Catalog Number |
1435 |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/21/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cdc adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.The reported event could not be confirmed; the controller dc adapter operated as expected during bench testing.Analysis of the device revealed that the device met specifications; the device passed visual examination and functional testing.The reported event could not be duplicated at the bench level.Per the instructions for use (ifu): the instructions for use and patient manual warn that disconnecting the controller from the driveline and disconnecting both power sources (batteries and ac or dc adapter) at the same time will stop the pump.At least one power source must be connected at all times.Additionally the steps and sequence for exchange of batteries and controllers are outlined.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Heartware will submit a supplemental report when new facts arises which materially alters information submitted in a previous mdr report.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.
|
| |
|
Event or Problem Description
|
|
It was reported by the coordinator that the patient complained of dc adapter not working.The device was exchanged with no reported patient consequences.
|
| |
|
Search Alerts/Recalls
|