Model Number MMT-1715K
Device Problems No Display/Image (1183); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had a partial display.The customer¿s blood glucose level was 96 mg/dl at the time of the incident.The customer reported that the insulin pump dropped a day before and now the screen was fading out.The customer was assisted with troubleshooting and the display did not return.Troubleshoot was performed for partial display.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Pump received with missing segments and partial display with vertical/horizontal black lines due to cracked lcd glass.Pump passed sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test and displacement test.Pump failed self test due to missing segments and partial display.Pump also received with scratched case, pillowing keypad overlay, and minor scratched lcd window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key6863687
MDR Text Key86273588
Report Number2032227-2017-47924
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169656840
UDI-Public(01)00643169656840
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG1K7Q2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/04/2017
Initial Date FDA Received09/13/2017
Supplement Dates Manufacturer Received01/22/2018
Supplement Dates FDA Received01/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age58 YR
Patient Weight236
-
-