• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN (IRVINE) ATK TURBOHAWK SMOOTH; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN (IRVINE) ATK TURBOHAWK SMOOTH; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number TH-LX-M
Medical Device Problem Codes Retraction Problem (1536); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: the turbohawk was inspected and observed the distal assembly showed the laser drilled coils bent/compressed at approximately 1.5 and 6cm from the cutter window.The cutter assembly was visible within the cutter window.The torque shaft was bent at an approximate 45 degree angle beneath the strain relief.The cutter driver was powered on and the thumb-switch was retracted.The cutter assembly retracted back into the cutter window as intended.The thumb-switch was pushed forward, however resistance was encountered.The cutter head was able to advance approximately 0.3cm from the cutter window.It was observed the outer diameter of the laser drilled tip was bent inward.The edge of the cutter head was contacting the inner diameter of the distal assembly at the location of the observed bend.The cutter was unable to advance within the distal assembly due to a blocked cutter lumen from a bend inwards of the laser drilled tip.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The physician was using a turbohawk directional atherectomy device to treat a moderately tortuous and moderately calcified mid left sfa lesion exhibiting 75% stenosis and plaque.The device was removed from the packaging and inspected with no issues noted.The device was prepped per the ifu.It was reported that the physician successfully inserted the device into the patient, made six passes, then removed and cleaned the device before re-advancing the device through the sheath and making several more passes through the lesion.The physician reported that they were unable to turn the device on and off without the thumbswitch sticking while at times, being used in the patient.It was reported that the device was sticking while the cutter was both inside and outside the housing.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATK TURBOHAWK SMOOTH
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
COVIDIEN (IRVINE)
9775 toledo way
irvine CA 92618
MDR Report Key6863940
Report Number2183870-2017-00391
Device Sequence Number18996514
Product Code MCW
UDI-Device Identifier00821684065754
UDI-Public00821684065754
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K061188
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2019
Device Catalogue NumberTH-LX-M
Device Lot NumberA324231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2017
Initial Report FDA Received Date09/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-