• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE; MAGIC 3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE; MAGIC 3 Back to Search Results
Model Number 53618G
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the patient attempted to insert the catheter, but was unable to due to a lack of lubricity.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿intermittent catheter with sure-grip sleeve hydrophilic coated silicone.The catheter is intended for urinary bladder drainage in patients requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/precautions and adverse reactions.Wash your hands thoroughly with soap and water.Release the sterile water from the foil packet.Tip the catheter pouch end-to-end three to six times so the water moves back and forth to thoroughly wet the catheter surface.Peel open the pack at the funnel end just enough to expose the insertion sleeve.Don¿t remove the catheter yet.Use the adhesive tab at the funnel end of the pack to stick the pack to a nearby vertical surface while preparing to catheterize.Wash the area around the meatus before catheterizing.Wash your hands again.Hold the insertion sleeve with your dominant hand and squeeze it to grip the catheter shaft as you remove the catheter from the pack.Next, hold the catheter funnel above the insertion sleeve with your other hand and slide the insertion sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.Using the insertion sleeve to hold the catheter firmly, gently pass the tip of the catheter into your urethra until the insertion sleeve nears the meatus.Repeat until urine starts to flow.Try to keep the catheter steady until urine stops flowing.When urine stops flowing, slowly withdraw the catheter, stopping if flow starts again, until the last few drops have drained.Finish by disposing of the catheter and its packaging.Wash your hands with soap and water.Warning: this is a single use device.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.Urethral catheter for urological use only.Discard after use.Made of silicone elastomer." (b)(4).
 
Event or Problem Description
It was reported that the patient attempted to insert the catheter, but was unable to due to a lack of lubricity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE
Common Device Name
MAGIC 3
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6864653
Report Number1018233-2017-04796
Device Sequence Number10942353
Product Code KOD
UDI-Device Identifier00801741069796
UDI-Public(01)00801741069796
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/28/2020
Device Model Number53618G
Device Catalogue Number53618G
Device Lot NumberJUBT0924
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/21/2017
Supplement Date Received by Manufacturer01/02/2018
Initial Report FDA Received Date09/14/2017
Supplement Report FDA Received Date01/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-