| Catalog Number |
180733-1 |
| Medical Device Problem Codes |
Device Operates Differently Than Expected (2913); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Surgeon attempted to put the poly in and it would not lock.After examining it, the base of it was not flat and therefore wouldn't lock into the base plate.Surgical delay of 5 to 7 minutes.Case type left pka.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding alleged seating/locking issue involving a mako insert was reported.The event was not confirmed.The device was returned in used condition.The scratches and gouges were observed throughout the device.Examination of the device by the material analysis engineer indicated that damage observed consistent with implantation process.-medical records received and evaluation: not performed as no medical records were provided.Device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies there has been no other event for the lot referenced.The event reported for base of the insert was not flat and therefore wouldn¿t lock into the base plate.The event was not confirmed due to no functional or dimensional inspection conducted since the device found damaged.The returned device showed scratches and gouges throughout the device.Examination of the device by the material analysis engineer indicated that damage observed consistent with implantation process.No further investigation for this event is required at this time.If the additional information are received, this investigation will be reopened and re-evaluated.
|
| |
|
Event or Problem Description
|
|
Surgeon attempted to put the poly in and it would not lock.After examining it, the base of it was not flat and therefore wouldn¿t lock into the base plate.Surgical delay of 5 to 7 minutes.Case type left pka.
|
| |
|