Intuitive surgical, inc.(isi) received the instrument associated with this complaint and has completed its evaluation.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was missing from clevis 0.032 x 0.046 in the clevis.The clevis did not exhibit any damage or wear marks.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to the reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
|
It was reported that during a da vinci-assisted prostatectomy procedure, the customer noted a broken off wire on the prograsp forceps instrument.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
|