Catalog Number 72200616SR
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/24/2017
Event Type  malfunction  
Event Description
It was reported the powermax elite mdu hand control buttons are stuck.No patient injury or complications were reported.
 
Manufacturer Narrative
The device was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and no issues were noted.Product failed functional testing with hand piece error when inserted into mdu receptacle on test control unit.Cause of hand piece errors is a corroded reverse and oscillate button springs that was stuck in the ¿on¿ position.Photos of button assembly corrosion attached.The complaint was confirmed and the root cause has been determined to be corrosion of the reverse and oscillate button assemblies.Factors that could have contributed to the event include a buildup of corrosion/debris around the button assemblies from cleaning chemical fluid ingression over time.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL MDU CNTRL PWRMX ELITE
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 w. william cannon drive
austin, TX 78735
MDR Report Key6865911
MDR Text Key86301670
Report Number1643264-2017-00539
Device Sequence Number1
Product Code HAB
UDI-Device Identifier00885556582305
UDI-Public(01)00885556582305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-