Model Number 420194-10
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument associated with this complaint and completed its evaluation.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to the reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
 
Event or Problem Description
It was reported that during a da vinci-assisted hysterectomy procedure, the customer noted loose wires on the mega needle driver instrument.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGA NEEDLE DRIVER
Common Device Name
ENDOSCOPIC INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
MDR Report Key6866404
Report Number2955842-2017-00604
Device Sequence Number125587
Product Code NAY
UDI-Device Identifier00886874111611
UDI-Public(01)00886874111611(10)N10170530
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420194-10
Device Lot NumberN10170530 375
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/06/2017
Initial Report FDA Received Date09/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-