| Catalog Number |
1435 |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/06/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One cdc adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.The reported event was confirmed via functional testing.Analysis of the device revealed that the device failed to meet specifications.The device passed visual examination but failed functional testing due to a broken connection within the cdc adapter power plug.The most likely root cause of the reported event may be related to a deficiency in the manufacturing process.Per the instructions for use (ifu): the instructions for use (ifu) and patient manual include a reference guide for both visual and tone alarms including potential causes and actions to take.The controller requires two power sources for safety: either two batteries or one battery and an ac adapter or dc adapter.Proper connection is verified when the ac/dc indicator on the controller turns green and the corresponding battery indicator turns off.If the ac/dc indicator doesn't turn green, the controller is using battery power and the "power disconnect" alarm will sound.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.
|
| |
|
Event or Problem Description
|
|
It was reported by the vad technician that the dc adapter would stop providing power to the controller a little while after being plugged in. the adapter was removed from service with no consequence to the patient. no further information was provided.
|
| |
|
Search Alerts/Recalls
|