MAUDE Adverse Event Report: HEARTWARE, INC HEARTWARE® VENTRICULAR ASSIST SYSTEM; CIRCULATORY ASSIST SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC HEARTWARE® VENTRICULAR ASSIST SYSTEM; CIRCULATORY ASSIST SYSTEM Back to Search Results
Catalog Number 1435
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/06/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One cdc adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.The reported event was confirmed via functional testing.Analysis of the device revealed that the device failed to meet specifications.The device passed visual examination but failed functional testing due to a broken connection within the cdc adapter power plug.The most likely root cause of the reported event may be related to a deficiency in the manufacturing process.Per the instructions for use (ifu): the instructions for use (ifu) and patient manual include a reference guide for both visual and tone alarms including potential causes and actions to take.The controller requires two power sources for safety: either two batteries or one battery and an ac adapter or dc adapter.Proper connection is verified when the ac/dc indicator on the controller turns green and the corresponding battery indicator turns off.If the ac/dc indicator doesn't turn green, the controller is using battery power and the "power disconnect" alarm will sound.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.
 
Event or Problem Description
It was reported by the vad technician that the dc adapter would stop providing power to the controller a little while after being plugged in. the adapter was removed from service with no consequence to the patient. no further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE® VENTRICULAR ASSIST SYSTEM
Common Device Name
CIRCULATORY ASSIST SYSTEM
Manufacturer (Section D)
HEARTWARE, INC
14400 nw 60th avenue
miami lakes FL 33014 3105
MDR Report Key6866415
Report Number3007042319-2017-03148
Device Sequence Number656697
Product Code DSQ
UDI-Device Identifier00888707000109
UDI-Public00888707000109
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2016
Device Catalogue Number1435
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2016
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/27/2017
Initial Report FDA Received Date09/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight137
-
-