| Model Number |
1884012HR |
| Medical Device Problem Code |
Stretched (1601)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The blade was received for analysis.Visually, the inner shaft was bent and a section of the inner spiral was included in a small bag which would have resulted in the reported event.The bend in the inner shaft was next to the distal face of the inner hub; the severity of the bend indicates that the device could not have been used if the defect were present prior to use; and therefore, it is likely the damage occurred during unloading of the blade from the handpiece or sometime after.The section of spiral wrap that was included in the bag measured approximately 0.24¿ long.Initially it could not be determined how the portion of the spiral wrap became detached because it should have been contained by the outer assembly; further analysis found that the outer cutter tip was not attached and slid off the end which would have exposed the spiral wrap.The detached portion of the tip measured approximately 0.44¿ long.When viewed under magnification there was damage to the teeth that is consistent with impact.There was no allegation of a defect prior to use.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician reported that while using the rad 12 blade the inner cylinder of the device was broken.It was confirmed that after st retching, a fragment of the spiral wrap detached; however, it did not fall into the patient or require retrieval.A back-up blade was used to complete the procedure.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|