| Catalog Number |
229435120 |
| Medical Device Problem Codes |
Loose or Intermittent Connection (1371); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
08/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that tibial trail does not lock onto tibial handle while the scrubtech was trying to piece the trials together.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: examination of the returned device confirms the reported event.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|