C.R. BARD, INC. (COVINGTON) -1018233 PERSONAL CATHETER¿ INTERMITTENT CATHETER; INTERMITTENT HYDROSIL CATHETER Back to Search Results
Model Number 61612
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that there were three issues concerning numerous catheters that the patient received.There was no water sachet, however the catheter was included.The sachet was cut and the catheter had a sharp tip, which caused bleeding.No medical intervention was required.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.One unopened samples were returned for evaluation.The sample was visually inspected for printing information in the package, completed package sealing without openings and free of damages (holes, perforations, tears, pleats or channels).During the visual inspection it was noted that the water pack was cut in a corner.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "for urological use only.Intended for use by patients for bladder management including urine drainage, collection and measurement.The devices are passed to the urinary bladder via the urethra.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/precautions and adverse reactions.This is a single use device.Do not resterilize.Do not reuse.Reuse of a single use device increases the risk of catheter acquired urinary tract infections.Wash your hands thoroughly with soap and water.Prior to opening the sealed catheter pouch, apply pressure to the foil packet to release the water.Ensure all water is released from the foil packet.Wet the catheter by holding the package with the printed side up and tip the package end-to-end three to six times to wet the catheter.This movement is required so that the water transfers back and forth over the catheter to fully wet the hydrophilic coating.Peel open the pack at the funnel/handle end just enough to expose 5 cm (2¿) of the catheter, or for products with an insertion sleeve, expose the insertion sleeve.Don¿t remove the catheter yet.Use the adhesive tab at the funnel end of the pack to stick the pack to a nearby vertical surface while preparing to catheterize.Wash the area around the meatus before catheterizing.Wash your hands again.Using the catheter with insertion sleeve (gripper): hold the insertion sleeve with your dominant hand and squeeze it to grip the catheter shaft as you remove the catheter from the pack.Next, hold the catheter funnel above the insertion sleeve with your other hand and slide the insertion sleeve down the shaft, stopping at about 6¿ from the tip.Release the funnel.Using the insertion sleeve to hold the catheter firmly, gently pass the tip of the catheter into your urethra until the insertion sleeve nears the meatus.Advance the catheter until urine starts to flow.Without insertion sleeve (gripper):" (b)(4).
 
Event or Problem Description
It was reported that there were three issues concerning numerous catheters that the patient received.There was no water sachet, however the catheter was included.The sachet was cut and the catheter had a sharp tip, which caused bleeding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONAL CATHETER¿ INTERMITTENT CATHETER
Common Device Name
INTERMITTENT HYDROSIL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6868318
Report Number1018233-2017-04818
Device Sequence Number496389
Product Code KOD
UDI-Device Identifier00801741069031
UDI-Public(01)00801741069031
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/28/2021
Device Model Number61612
Device Catalogue Number63612
Device Lot NumberJUAX0686
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2017
Initial Date Received by Manufacturer 08/22/2017
Supplement Date Received by Manufacturer01/25/2018
Initial Report FDA Received Date09/15/2017
Supplement Report FDA Received Date01/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-