BOSTON SCIENTIFIC - SPENCER EXCELON¿; BRONCHOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Model Number M00564111
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Visual evaluation of the device revealed that the sheath and needle were bent approximately 11cm from the distal end.In addition the inner and outer sheath were detached.Functional analysis could not be performed due to the condition of the device.Based on the evaluation of the returned device, these failures are likely due to anatomical or procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) was performed and no deviations were found.A search of the complaint database confirmed that no similar complaints exist for the specified batch.
 
Event or Problem Description
It was reported to boston scientific corporation that an excelontransbronchial aspiration needle was used in the main carina during a transbronchial needle aspiration (tbna) procedure performed on (b)(6) 2017.According to the complainant, during the procedure the needle was stuck on distal stop.The procedure was completed with this device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.The investigation found that the needle was bent, this is now deemed an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCELON¿
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key6868393
Report Number3005099803-2017-02744
Device Sequence Number461626
Product Code EOQ
UDI-Device Identifier08714729456469
UDI-Public(01)08714729456469(17)20190131(10)18860143
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K040018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date01/31/2019
Device Model NumberM00564111
Device Catalogue Number6411
Device Lot Number18860143
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/22/2017
Initial Report FDA Received Date09/15/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age46 YR
-
-