Model Number 60-02-60
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/23/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to reproduce the reported issue and identified the root cause to be a configuration issue.The customer owns two cp5 units and the way there were configured was causing a conflict.The service representative flashed the second flow module to be a second device to correct the conflict.No further issues were reported.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Evaluated on site by livanova technician.
 
Event Description
Livanova (b)(4) received a report that the flow value on the display of the centrifugal pump 5 (cp5) was erratic and bounced up and down during a procedure, though the user was not touching the flow knob.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5 (CP5)
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6869111
MDR Text Key86429857
Report Number9611109-2017-00722
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817901006
UDI-Public010403381790100611170705
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 09/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-60
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-