Model Number 4415 |
Device Problem
Device Reprocessing Problem (1091)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/11/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This complaint was found when the field service representative (fsr) was servicing the device for related complaint (b)(4), mdr #1828100-2017-00197.The fsr cleaned the inline filter due to large amounts of calcium deposits.Brown/yellow water was noted when refilling the unit.During testing the inline filter was filled with calcium deposits again.The customer was advised to follow the instructions for use (ifu) as cleaning the unit for clinical use is the customer's responsibility.
|
|
Event Description
|
It was reported that during the use of the device for a non-clinical activity, the water from the heater cooler was discolored.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The field service representative (fsr) cleaned the pressure switch bushing and fixed the problem.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|