| Model Number |
48811234 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that locking mechanism was damaged during screw removal.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection; device history review; complaint history review; risk assessment; results: visual inspection confirmed the spring bar was deformed.Manufacturing files were reviewed and no anomalies were found.Conclusion: the possible root causes include:screw not fully seated below spring bar, poor visibility of spring-bar, spring-bar damaged prior to use, blocker locked before visual confirmation of screw below spring bar.
|
| |
|
Event or Problem Description
|
|
It was reported that locking mechanism was damaged during screw removal.
|
| |
|