| Model Number |
MMT-523NAP |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump alarmed button error.The customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump alarmed button error and no significant event leading to button error.Button error troubleshooting was performed and confirmed that time is not advancing even after the battery has ben changed.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Search Alerts/Recalls
|