U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK SPECTRUM; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK SPECTRUM; FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number G44128
Date of Event 01/19/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Brand name- cook spectrum minocycline/rifampin impregnated triple lumen central venous catheter tray.Rpn- c-utlmy-701j-rsc-abrm-hc-fst-a-rd.(b)(4).This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
During an attempt to insert a central venous catheter into the femoral vein, difficulty was encountered with the wire kinking, bending and coiling, inhibiting the passage of the catheter.The procedure was being performed on an elderly male patient that was in shock, requiring vasopressors.The procedure was performed by a pulmonary fellow under the supervision of an attending physician.Ultrasound of the area did not indicate any deviations in the patient¿s anatomy both proximal and distal to the insertion site.The needle was inserted without issue under ultrasound guidance; the wire was advanced to the lower 1/3 of the superior vena cava.The needle was removed and a small skin nick was made on both sides of the wire.The dilator was inserted over the wire and it is unclear if there was any issue inserting the dilator.At some point in the dilation, the fellow described the wire as going slack or losing rigidity.The dilator was removed and the catheter was inserted over the wire.At this point the catheter did not advance as expected.The fellow removed the catheter and the wire.It was noted that there was a "huge" kink in the proximal end of the wire approximately at the distance of the tip of the dilator.The fellow opened up another manufacturer¿s cvc kit and performed a femoral central line insertion on the contralateral side.There were no issues with this insertion.The physician described the event as rare.It was also mentioned that the guidewires were kinking, bending and coiling and they are unable to advance the catheter over the guide wire.¿multiple failed attempts at placing femoral central and arterial lines by vastly experienced specialists in the micu¿.Specific details were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SPECTRUM
Common Device Name
FOZ CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6872132
Report Number1820334-2017-03077
Device Sequence Number4204462
Product Code FOZ
UDI-Device Identifier00827002441286
UDI-Public(01)00827002441286(17)190401(10)7354173
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberG44128
Device Catalogue NumberC-UTLMY-701J-RSC-ABRM-HC-FST-A-RD
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2017
Initial Report FDA Received Date09/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
-
-