| Model Number |
8886605731 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter: during a tapp procedure, the package of the first packet was opened, however the needle and suture thread was disengaged.Replaced by a new packet, however the needle and suture thread was disengaged easily.The event occurred during the procedure but the product was not used for patient.The status of the patient: no problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned products noted one suture and one needle.The suture was detached from the needle.It appeared that the needle had disengaged from the suture under tension.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|