|
The customer called via phone call that his pump alarmed compromised forced sensor.At the time of the incident, his blood glucose 168 mg/dl.The customer said that he was in the middle of trying to change his infusion set when the alarm occurred and that he is stuck in preparing prime loop.During prime rewind troubleshooting, the customer said that insulin was squirting out and continued to drip after the motor stopped during the manual prime process.He said that the pump is alarming compromised forced sensor and that they are unable to exit the preparing to prime loop.He said that he is stuck in rewind complete/bolus delivery loop.The customer was advised to hold the act button down until he received the 2nd series of beeps and/or numbers appeared on his display.He said that nothing happened when he held the button down.The compromised forced sensor alarm recurred.He was advised to discontinue use of the insulin pump and to revert to the back-up plan per his doctor¿s orders.The insulin pump will be returning for analysis.
|
|
Insulin pump passed the displacement test but compromised force sensor system alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform the occlusion, prime/compromised force sensor system and excessive no delivery test or verify prime/fill anomaly due to compromised force sensor system alarm.No moisture damage on the lcd board, mother board, interface board, rf board, motor, vibrator motor and battery tube assembly during visual inspection.Insulin pump was received with cracked case at the display window corner, broken reservoir tube lip, minor scratched lcd window, missing end cap sticker, stained address/serial number label, cracked belt clip slot, cracked case reservoir tube window corner, cracked lcd window and cracked battery tube threads.
|