| Model Number |
CB006 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 13-sep-2017 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
Fill volume: 400 ml, flow rate: 14ml initially, then 8ml overnight, procedure: total knee arthroplasty (tka), cathplace: abductor nerve block.It was reported that patient a experienced a fast flow due to a red tab not being pulled out from the bolus button.The entire volume infused overnight.The patient was reported to be fine.No further information was provided.
|
| |
|