| Model Number |
MMT-7008B |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 121 mg/dl and the sensor glucose was 46 mg/dl at the time of the incident.The customer stated he had false readings and also shutoff in the middle of the night suspended.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The product will be replaced.The product will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened/used sensor and performed continuity resistance test and sensor passed.Performed bicarbonate buffer test and sensor passed per specifications.
|
| |
|