Model Number U41502H22
Medical Device Problem Codes Break (1069); Difficult to Open or Remove Packaging Material (2922)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that upon opening the package, in preparation for the angioplasty procedure, the balloon stylet allegedly could not be removed.It was further reported that another pta catheter was used to perform the procedure.There was no patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Conclusion:the device was returned.A visual inspection found a complete break in the outer catheter, as well as stretching of the inner guidewire lumen.Additionally, the balloon guard was found to be bunched.Based on the observed damage to the device and balloon guard, the investigation is confirmed for the reported difficultly removing the device from the balloon stylet and guard.Based on the condition of the returned sample, the difficulty experienced during removal led to excessive force being applied, resulting in the break in the outer catheter and other damage to the device.However, the definitive root cause for the difficulty removing the stylet and guard from the device could not be determined based upon available information.Labeling review: the current ifu (instructions for use) states: dilatation catheter preparation: remove the balloon guard and stylet by grasping the balloon catheter just proximal to the balloon and with the other hand, gently grasp the balloon protector and slide distally off of the balloon catheter.
 
Event or Problem Description
It was reported that upon opening the package, in preparation for the angioplasty procedure, the balloon stylet allegedly could not be removed.It was further reported that another pta catheter was used to perform the procedure.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 014 PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6874596
Report Number2020394-2017-01169
Device Sequence Number13595904
Product Code LIT
UDI-Device Identifier00801741053658
UDI-Public(01)00801741053658
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121856
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberU41502H22
Device Catalogue NumberU41502H22
Device Lot NumberCMAY0552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2017
Initial Date Received by Manufacturer 08/22/2017
Supplement Date Received by Manufacturer10/09/2017
Initial Report FDA Received Date09/18/2017
Supplement Report FDA Received Date10/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-