| Model Number |
126037 |
| Medical Device Problem Code |
Deflation Problem (1149)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that prior to use on a patient, while checking the cuff, the cuff did not deflate.The customer reported that there was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample was returned and visual examination showed that the shape returned unit had been altered using the stylet wire.The stylet wire was removed and inflation/deflation testing was successful.There was no fault found with the device.The probable root cause is that the cuff test was carried out when the shape of the tube was incorrectly altered with the stylet wire.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that prior to use on a patient, while checking the cuff, the cuff did not deflate.The customer reported that there was no patient involvement.
|
| |
|