ARTHREX INC. UNIVERSAL GLENOID - BASEPLATE SMALL; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
AR-9120-01 |
| Health Effect - Clinical Code |
Fall (1848)
|
| Date of Event |
09/01/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient demographics (date of birth, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.This is one of two submissions from the same event.The others are cc123986-line 205050-00357.The device is expected but has not yet been received.As reported in the event, the patient supposedly fell which dislodged the baseplate/glenosphere.The most likely cause of this event is failure to follow the post-op rehab protocol.The directions for use states: post-operatively, until healing is complete the fixation provided by this device should be protected.The post-operative regimen prescribed by the physician should be strictly followed to avoid adverse stresses applied to the implant.This is the first complaint of this type for this part/lot combination.The potential causes of this event are being communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had reverse total shoulder on (b)(6) 2016.The patient supposedly fell in (b)(6) which dislodged the baseplate/glenosphere.Revision surgery was completed on (b)(6) 2017, converting to hemi.**additional information received on 9/5/2017: the devices explanted during revision: ar-9120-01 lot 160012712 (line 204892), ar-9165-15nl lot 1444855 (line 205048), ar-9145-30 lot 160024110 (line 205049), ar-9504s-04 lot 160036711 (line 205050), ar-9503s-06 lot 160010209 (line 205051), ar-9145-30 lot 160009610 (line 205056).The devices implanted during the revision were ar-9950-19rca and ar-9502-36arca.Quality of bone was reported to be poor.Male, (b)(6).
|
| |
|
Search Alerts/Recalls
|
|
|