• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER; MAGIC 3 CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER; MAGIC 3 CATHETER Back to Search Results
Model Number 51514
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter lacked lubricity.
 
Additional Manufacturer Narrative
Received 5 opened magic3 intermittent catheter lot samples.The reported event was inconclusive due to the poor sample condition of all samples received.The samples were visually inspected for printing information in the package, the package sealing was without openings and free of damages (holes, perforations, tears, pleats or channels).During the visual inspection, the samples were found to be already opened and used.A functional evaluation could not be done due to the sample conditions.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "the catheter is intended for urinary bladder drainage in adult males and females requiring catheterization for management of incontinence, voiding dysfunction, and surgical procedures.Efficacy of the catheter in preventing urinary tract infection during intermittent use has not been established.The device is not intended to be used as a treatment for active urinary tract infection.Please contact your physician to determine which product options are best for you, paying close attention to product warnings/precautions and adverse reactions.Instructions for use the catheter becomes slippery when wetted with water, eliminating the need for a separate lubricant.For your added convenience, this catheter is packaged with its own sterile water.Simply release the water from its foil packet and then tip the un-opened catheter package end-to-end.The catheter acts like a magnet to attract the water and activate its slippery coating.Follow these steps for best results.Release the water prior to opening the sealed catheter pouch: apply pressure to the foil packet to release the water.Ensure all water is released from the foil packet.Wet the catheter hold package with printed side up.Tip package end-to-end three to six times to wet catheter.This movement is required so that the water transfers back and forth over the catheter to fully wet the hydrophilic coating.Use the catheter peel back package to expose funnel end of catheter.If desired, use the self-adhesive tape on the package to temporarily attach the pouch to any dry vertical surface.Remove catheter and use according to physician¿s instructions.Active ingredient the catheter contains 10.2 +/- 2.0 micrograms nitrofurazone (5 nitro-2-furaldehyde semicarbazone) per mm2.The amount of nitrofurazone per mm2 remains constant regardless of catheter french size.The funnel end of the catheter is excluded in calculating the catheter surface area." (b)(4).
 
Event or Problem Description
It was reported that the catheter lacked lubricity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER
Common Device Name
MAGIC 3 CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key6878810
Report Number1018233-2017-04890
Device Sequence Number800686
Product Code MJC
UDI-Device Identifier00801741068188
UDI-Public(01)00801741068188
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K033477
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/28/2022
Device Model Number51514
Device Catalogue Number51514
Device Lot NumberJUBQ1393
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2017
Initial Date Received by Manufacturer 08/29/2017
Supplement Date Received by Manufacturer10/31/2017
Initial Report FDA Received Date09/20/2017
Supplement Report FDA Received Date11/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-