• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM Back to Search Results
Model Number MT22495-BLU
Medical Device Problem Codes No Audible Alarm (1019); Use of Device Problem (1670)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had no audio output.No additional event or patient information is available.No product or data were provided for evaluation.The reported no audio output could not be confirmed.A root cause could not be determined.Reportedly, a pediatric patient was using an adult receiver.Labeling indicates: the dexcom g4 system is not approved for use in children or adolescents, pregnant women or persons on dialysis.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The receiver was returned for evaluation.The device was visually inspected and no defects were found.Receiver was charged and booted up.A manual test was performed and speaker did not sound.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM DEXCOM SHARE SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key6879934
Report Number3004753838-2017-74250
Device Sequence Number1324976
Product Code MDS
UDI-Device Identifier20386270000174
UDI-Public20386270000174
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22495-BLU
Device Catalogue NumberSTK-RR-BLU
Device Lot Number5202728
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2017
Supplement Date Received by Manufacturer09/28/2017
Initial Report FDA Received Date09/20/2017
Supplement Report FDA Received Date10/23/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3004753838-02/29/16-001C
Patient Sequence Number1
Patient Age8 YR
Patient Weight32
-
-