| Model Number |
MT22495-BLU |
| Medical Device Problem Codes |
No Audible Alarm (1019); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2017 that on (b)(6) 2017 the receiver had no audio output.No additional event or patient information is available.No product or data were provided for evaluation.The reported no audio output could not be confirmed.A root cause could not be determined.Reportedly, a pediatric patient was using an adult receiver.Labeling indicates: the dexcom g4 system is not approved for use in children or adolescents, pregnant women or persons on dialysis.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.The device was visually inspected and no defects were found.Receiver was charged and booted up.A manual test was performed and speaker did not sound.The reported event of no audio output was confirmed.The root cause was determined to be a defective speaker.
|
| |
|
Search Alerts/Recalls
|