Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned.The investigation was confirmed for catheter twisting, a hole in the catheter and material separation of the catheter and metal distal tip.The investigation wa inconclusive for guidewire incompatibility as patency testing was unable to be performed due to sample condition.The root cause could not be determined based upon available information.It was unknown if procedural factors contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings and precautions: - when using the crosser® catheter in tortuous anatomy, the use of a support catheter is recommended to prevent kinking or prolapse of the crosser® catheter tip.Kinking or prolapse of the tip could cause catheter breakage and/or malfunction.-do not exceed 5 minutes of activation time as crosser catheter malfunction may occur.If 5 minutes of activation time is achieved exchange for a second crosser catheter before resetting the crosser generator.Adverse events: - as with most percutaneous interventions, potential adverse effects include: bleeding which may require transfusion or surgical intervention, hematoma, perforation, dissection, guidewire entrapment and/or fracture, hypertension/hypotension, infection or fever, allergic reaction, pseudoaneurysm or fistula aneurysm, acute reclosure, thrombosis, ischemic events, distal embolization, excessive contrast load resulting in renal insufficiency or failure, excessive exposure to radiation, stroke/cva, restenosis, repeat catheterization / angioplasty, peripheral artery bypass, amputation, death or other bleeding complications at access site.Set up: - back the rigid portion of the catheter out of the end of the hoop, then carefully unsnap the catheter from the hoop with a gentle twisting motion.- set the irrigation system to 0.3ml/sec (18ml/min) and switch irrigation system "on".Allow irrigation to flow for approximately 10-15 seconds to purge all air from the crosser catheter irrigation lumen.Interventional use: - advance the guidewire and crosser catheter 14p, 14s, or 18 to the lesion site.Withdraw the guidewire approximately 1cm within the catheter so that the tip of the crosser catheter is the leading edge.Slowly advance the catheter tip through the lesion.Apply steady, constant pressure so the tip of the catheter is engaged to the lesion.- upon successful recanalization of the lesion, advance the guidewire distal to the lesion and then withdraw the crosser 14s catheter.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|