• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETER; CATHETER FOR CROSSING TOTAL OCCLUSIONS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CROSSER CTO RECANALIZATION CATHETER; CATHETER FOR CROSSING TOTAL OCCLUSIONS Back to Search Results
Catalog Number CRU14SA
Medical Device Problem Codes Hole In Material (1293); Material Separation (1562); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/21/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned.The investigation was confirmed for catheter twisting, a hole in the catheter and material separation of the catheter and metal distal tip.The investigation wa inconclusive for guidewire incompatibility as patency testing was unable to be performed due to sample condition.The root cause could not be determined based upon available information.It was unknown if procedural factors contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings and precautions: - when using the crosser® catheter in tortuous anatomy, the use of a support catheter is recommended to prevent kinking or prolapse of the crosser® catheter tip.Kinking or prolapse of the tip could cause catheter breakage and/or malfunction.-do not exceed 5 minutes of activation time as crosser catheter malfunction may occur.If 5 minutes of activation time is achieved exchange for a second crosser catheter before resetting the crosser generator.Adverse events: - as with most percutaneous interventions, potential adverse effects include: bleeding which may require transfusion or surgical intervention, hematoma, perforation, dissection, guidewire entrapment and/or fracture, hypertension/hypotension, infection or fever, allergic reaction, pseudoaneurysm or fistula aneurysm, acute reclosure, thrombosis, ischemic events, distal embolization, excessive contrast load resulting in renal insufficiency or failure, excessive exposure to radiation, stroke/cva, restenosis, repeat catheterization / angioplasty, peripheral artery bypass, amputation, death or other bleeding complications at access site.Set up: - back the rigid portion of the catheter out of the end of the hoop, then carefully unsnap the catheter from the hoop with a gentle twisting motion.- set the irrigation system to 0.3ml/sec (18ml/min) and switch irrigation system "on".Allow irrigation to flow for approximately 10-15 seconds to purge all air from the crosser catheter irrigation lumen.Interventional use: - advance the guidewire and crosser catheter 14p, 14s, or 18 to the lesion site.Withdraw the guidewire approximately 1cm within the catheter so that the tip of the crosser catheter is the leading edge.Slowly advance the catheter tip through the lesion.Apply steady, constant pressure so the tip of the catheter is engaged to the lesion.- upon successful recanalization of the lesion, advance the guidewire distal to the lesion and then withdraw the crosser 14s catheter.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the recanalization catheter allegedly would not load onto the guidewire, rendering the device unusable.Reportedly, another guidewire and balloon catheter were used to complete the procedure.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETER
Common Device Name
CATHETER FOR CROSSING TOTAL OCCLUSIONS
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6879938
Report Number2020394-2017-01179
Device Sequence Number633551
Product Code PDU
UDI-Device Identifier10801741125567
UDI-Public(01)10801741125567
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/28/2018
Device Catalogue NumberCRU14SA
Device Lot NumberGFAY0448
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2017
Initial Report FDA Received Date09/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-